Beam calculator

Ultrasound beam and resolution

Near field, beam width, depth of field, axial resolution, penetration and grating lobes, updated as you type.

Beam

Beam sketch

Pulse and axial resolution

Penetration

Array

On-axis pressure, unfocused circular piston of the same diameter |p(z)| ∝ |sin(π/λ·(√(z² + a²) − z))|, a = D/2
How this is calculated
  • Lengths in mm, time in µs, frequency in MHz; c in m/s is c/1000 mm/µs. The wavelength is λ = c / f.
  • Near-field (Fresnel) length N = D² / (4λ), valid for D ≫ λ. The last on-axis maximum of a circular piston is at (D² − λ²) / (4λ). A lens or delays cannot focus a beam beyond about N, so when F ≥ N (or the aperture is unfocused) we evaluate the beam at its natural focus, z ≈ N.
  • Lateral −6 dB width at the focus is K·λ·F# with F# = F / D. K comes from the focal-plane pattern: the Airy pattern 2J₁(x)/x for a circular disc and sin(u)/u for a rectangular aperture in its imaging plane. One-way uses amplitude 0.5; pulse-echo squares the pattern, so it uses 0.707. Solving these gives K = 1.41 and 1.03 (circular), 1.21 and 0.89 (rectangular).
  • Depth of field (−6 dB focal zone length) ≈ 7·λ·F#²: a common approximation. The real focal zone is not symmetric about F.
  • For a Gaussian pulse, the −6 dB duration and the −6 dB bandwidth satisfy τ·Δf = 4 ln 2 / π ≈ 0.883, so τ ≈ 0.883 / (FBW·f). Spatial pulse length is c·τ and axial resolution is about half of it.
  • Round-trip attenuation 2·α·f·d assumes attenuation rises linearly with frequency (water and many solids rise closer to f²). The maximum useful depth is where it uses up the dynamic range: d_max = DR / (2·α·f). This ignores reflection strength, focusing gain and noise.
  • Grating lobes of an array appear at sin θg = sin θs ± λ/p whenever |sin θg| ≤ 1. There are none up to a steering angle θ if p ≤ λ / (1 + sin θ), and none at any angle if p ≤ λ/2.
  • The beam sketch is to scale in depth; the envelope is a smooth approximation of the one-way −6 dB beam, not a field simulation.

An approximate, linear, textbook model for design intuition. It is not for safety or dosimetry decisions. Nothing you enter leaves your browser.